Description / Abstract:
This Standard covers air data equipment (hereinafter designated
the instrument) which when connected to sources of aircraft
electrical power, static pressure, total pressure and outside air
temperature (singly or in combination) provides some or all of the
following computed air data output signals:
Pressure Altitude*
Pressure Altitude (Reporting)
Baro-Corrected Pressure Altitude*
Vertical Speed*
Computed Airspeed*
Mach Number*
Maximum Allowable Airspeed*
Static Air Temperature (*when used as an alternate for total
temperature)
Total Temperature*
Altitude Rate
Overspeed Warning
Altitude Hold
Airspeed Hold
Mach Hold
qc (impact pressure)
True Airspeed
Others
Limits:
Full scale outputs marked with an asterisk may supply required
pilots flight instruments as specified in Federal Aviation
Regulations (e.g., Part 25, paragraph 25.1303) and when so used,
shall meet the minimum accuracy requirements of Paragraph 4 of this
document (AS 417A). Other outputs of the same function, signal
characteristic and similar usage such as flight recorder altitude
may have an added "correspondence tolerance" which shall not exceed
50% of the minimum standard tolerance unless otherwise specified by
existing Federal requirements.
Those outputs supplied to other subsystems for scheduling
purposes are not the concern of this minimum performance
specification but should be reviewed for compatibility in the
particular installation.
The above outputs shall include correction for certified static
source errors as provisioned and matched to the characteristics of
the particular type and model aircraft in which the instrument is
intended to be used. The tolerances expressed in Paragraph 4
include the allowance for static source error corrections up to ±
300 ft for any altitude-airspeed combination. Necessity for greater
error correction requires special review and approval of tolerances
based on the installation.
Those outputs intended for particular using subsystems and being
either of a discrete, incremental, or rate characteristics and not
covered by 2.1.1 above shall meet the specific requirements for the
respective category of output as may apply in Paragraph 4 with no
added correspondence allowance where accuracy is concerned for
duplicate outputs.
The instrument shall not be damaged by having applied to its
pressure fittings any of the following pressures singly or in
combination: Static Pressure Fitting - Approximately 1.4 inch Hg
Abs to 37 inch Hg Abs. Total Pressure Fitting - Approximately 0
inch Hg Abs to 50 inch Hg Abs, with the additional provision that
the instrument shall not be required to withstand an excess of
Total over Static Pressure of more than 19 inch Hg, nor an excess
of Static over Total Pressure of more than 2 inch Hg.
In the event that the above limits are likely to be exceeded,
Transducer over Pressure protection shall be provided by the
user.